Siemens 7ML19985AB01 instruction manual = head, cm = flow velocity, cm/sec

Models: 7ML19985AB01

1 112
Download 112 pages 47.88 Kb
Page 83
Image 83

TRAPEZOIDAL AREA X VELOCITY, P3 = 14

‘U’ parameters required *

‘U’ parameters calculated **

U0

= channel top width

B

U3 = area (h)

U1

= channel base width

b

 

U2

= channel depth

ht

 

*obtain from manufacturer’s specifications.

**calculated by OCM-3. May be viewed by accessing ‘U’ parameter.

Reference

ABSOLUTE CALCULATION, P4 = 0¤

For flows that can be calculated by the equation:

q = 1/1000 x (b + mh) x v

m = (B - b)/d

where : q = flow rate, l/sec

B = channel top width, cm

b = channel bottom width, cm d = depth of channel, cm

h = head, cm

v = flow velocity, cm/sec

RATIOMETRIC CALCULATION, P4 = 1¤

For flows that can be calculated by the equation:

q = qcal x A/Acal x v/vcal

A = (b + mh) x h

Acal = (b + mhcal) x hcal

m = (B - b)/d

where : q = flow rate, l/sec

B = channel top width, cm

b = channel bottom width, cm d = depth of channel, cm

h = head, cm

v = flow velocity, cm/sec

¤Refer to Operation \ Flow Calculation.