Owner's Record

The model and serial numbers are located at the rear. Record these numbers in the space provided below. Refer to these numbers whenever you call upon your Sony dealer regarding this product.

Model No. ____________________

Serial No. ____________________

WARNING

To reduce the risk of fire or electric shock, do not expose this apparatus to rain or moisture.

To avoid electrical shock, do not open the cabinet. Refer servicing to qualified personnel only.

THIS APPARATUS MUST BE EARTHED.

Symbols on the products

This symbol indicates the equipotential terminal which brings the various parts of a system to the same potential.

This symbol is intended to alert the user to the presence of important operating and maintenance (servicing) instructions in the literature accompanying the appliance.

Important safeguards/notices for use in the medical environments

1.All the equipments connected to this unit shall be certified according to Standard IEC60601-1, IEC60950, IEC60065 or other IEC/ISO Standards applicable to the equipments.

2.When this unit is used together with other equipment in the patient area*, the equipment shall be either powered by an isolation transformer or connected via an additional protective earth terminal to system ground unless it is certified according to Standard IEC60601-1.

*Patient Area

3.The leakage current could increase when connected to other equipment.

4.This equipment generates, uses, and can radiate radio frequency energy. If it is not installed and used in accordance with the instruction manual, it may cause interference to other equipment. If this unit causes interference (which can be determined by unplugging the power cord from the unit), try these measures: Relocate the unit with respect to the susceptible equipment. Plug this unit and the susceptible equipment into different branch circuit. Consult your dealer. (According to standard EN60601-1-2 and CISPR11, Class B, Group 1)

Caution

When you dispose of the unit or accessories, you must obey the law in the relative area or country and the regulation in the relative hospital.

WARNING on power connection

Use a proper power cord for your local power supply.

1.Use the approved Power Cord (3-core mains lead) / Appliance Connector / Plug with earthing-contacts that conforms to the safety regulations of each country if applicable.

2.Use the Power Cord (3-core main lead) / Appliance

Connector / Plug conforming to the proper ratings (Voltage, Ampere).

If you have questions on the use of the above Power Cord / Appliance Connector / Plug, please consult a qualified service personnel.

WARNING on power connection for medical use

Please use the following power supply cord.

With connectors (plug or female) and cord types other than those indicated in this table, use the power supply cord that is approved for use in your area.

 

United States

Canada

 

 

 

Plug Type

HOSPITAL GRADE*

HOSPITAL GRADE*

 

 

 

Female end

E62405, E35708

LR53182, LL022442,

 

 

LL088408

Cord type

E159216, E35496

LL112007-1, LL20262,

 

Min.Type SJT

LL32121, LL84494

 

Min.18 AWG

Min.Type SJT

 

 

Min.18AWG

 

 

 

Minimum cord set

10A/125V

10A/125V

rating

 

 

Safety approval

UL Listed

CSA

 

 

 

*Note: Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked ‘Hospital Only’ or ‘Hospital Grade’.

When installing the installation space must be secured in consideration of the ventilation and service operation.

Do not block the ventilation slots at the rear, and vents of fans.

Leave a space around the unit for ventilation.

Leave more than 10 cm of space in the rear of the unit to secure the operation area.

2