LOG SENSE6.Command Specification

6.6.6. Tape Alert Log Page

Bit

7

6

5

4

 

3

 

2

1

0

Byte

 

 

 

 

 

 

 

 

 

 

0

Reserved

 

Page Code (2Eh)

 

 

1

 

 

 

Reserved

 

 

 

 

 

2

(MSB)

 

 

Page Length (140h)

 

 

3

 

 

 

 

 

 

 

 

 

(LSB)

5n-1

(MSB)

 

 

Parameter Code (n)

 

 

5n

 

 

 

 

 

 

 

 

 

 

5n+1

DU (0)

DS (1)

TSD (0)

ETC (0)

 

 

TMC (00)

LBIN (0)

LP (0)

5n+2

 

 

 

Parameter

Length (1)

 

 

5n+3

 

 

 

Value of Flag

 

 

 

 

The value of n in the Parameter Code field shall be in the range of 1 to 64.

A value of one in the Value of Flag field indicates the flag is set.

Each flag will be cleared to zero in the following conditions:

At power on cycle

When Tape Alert Log page is read

When specified corrective action has been performed

SCSI bus reset or Bus Device Reset message

Log Select reset

There are three types of flags, listed below in order of increasing severity:

Table 6-20: Tape Alert flag types

Severity

Urgent User

Risk of Data Loss

Explanations

 

Intervention

 

 

C (Critical)

X

X

 

W (Warning)

 

X

X

I (Information)

 

 

X

SONY AIT-3Ex drive SDX-800V series Ver.1.1

6-19

Page 77
Image 77
Sony SDX-800V/R manual Tape Alert Log, Bit Byte Reserved Code 2Eh, Length 140h, Parameter Code n 5n+1