LOG SENSE

6.Command Specification

6.6.14. MIC Variable Length Information Page (Sony Unique)

Page 3Eh is a variable length page. That is not defined by ANSI. This page is Sony Vendor Unique.

Important NOTE:

1.Only one Parameter Code in Page 3Eh is transferred at a time.

2.Byte position 3 of Parameter Code in the Page 3Eh is Reserved. (Should be set to ZERO.)

3.Parameter Length for the contents of the Parameter Code in Page 3Eh is stored at the byte position 4 and 5.

4.The Page Length field in the Page Code indicates the total number of bytes that follow this byte. This value is independent of what is specified for the Allocation Length in the CDB.

Table 6-35: MIC Variable Length Information Page

Bit

7

6

5

4

 

3

2

1

0

Byte

 

 

 

 

 

 

 

 

 

0

 

 

 

Page Code (3Eh)

 

 

 

1

 

 

 

 

Reserved

 

 

 

2

(MSB)

 

 

Page Length (n-3)

 

 

 

3

 

 

 

 

 

 

 

 

(LSB)

Parameter Code

 

 

 

 

Description

 

 

 

0001h

 

User Volume Note

 

 

 

 

 

 

0002h

 

User Partition Note for Partition 0

 

 

 

 

 

 

 

 

 

 

 

 

 

0

 

(MSB)

 

 

 

Parameter Code (0001h)

 

 

1

 

 

 

 

 

(User Volume Note)

 

(LSB)

2

 

DU (0)

DS (1)

TSD (0)

ETC (0)

 

TMC (00)

LBIN (1)

LP (1)

3

 

 

 

 

 

Reserved

 

 

 

4

 

(MSB)

 

 

 

User Data Length

 

 

5

 

 

 

 

 

 

 

 

 

(LSB)

6

 

(MSB)

 

 

 

User Data

 

 

 

N

 

 

 

 

 

 

 

 

 

(LSB)

 

 

 

 

 

 

 

 

 

0

 

(MSB)

 

 

 

Parameter Code (0002h)

 

 

1

 

 

 

 

(User Partition Note for Partition #0)

 

(LSB)

2

 

DU (0)

DS (1)

TSD (0)

ETC (0)

 

TMC (00)

LBIN (1)

LP (1)

3

 

 

 

 

 

Reserved

 

 

 

4

 

(MSB)

 

 

 

User Data Length

 

 

5

 

 

 

 

 

 

 

 

 

(LSB)

6

 

(MSB)

 

 

 

User Data

 

 

 

N

 

 

 

 

 

 

 

 

 

(LSB)

Note:

1.If there is no Note data in MIC for the Parameter Code specified in the CDB a Check Condition status will be returned.

2.If the CDB allocation length is ZERO a Good status is always returned and there is no data phase.

3.If the CDB allocation length is non-ZERO and there is data in MIC for the Parameter Code data will be returned up to the size of the MIC data, as limited by the allocation length. For example, if the CDB allocation length is 0FFh and MIC data size for the Parameter Code is 020h 020h data will return. If the CDB allocation length is 010h and the MIC data size for the Parameter Code is 020h only 010h data will be returned.

SONY AIT-3Ex drive SDX-800V series Ver.1.1

6-37

Page 95
Image 95
Sony SDX-800V MIC Variable Length Information Page Sony Unique, Important Note, Length n-3, Parameter Code Description