Command Descriptions
2–100 TDS Family Oscilloscope Programmer Manual

 






 specifies the ASCII representation of signed integer () data. If
this is the value at power-on, the WFMPre values for BN_Fmt, BYT_Or, and
ENCdg are set as RP, MSB, and ASC respectively.
 specifies signed integer data-point representation with the most
significant byte transferred first. This format results in the fastest data transfer
rate when DATa:WIDth is set to 2.
The range is –128 to 127 when DATa:WIDth is 1. Zero is center screen. The
range is –32768 to 32767 when DATa:WIDth is 2. The upper limit is one
division above the top of the screen and the lower limit is one division below the
bottom of the screen.
 specifies positive integer data-point representation with the most
significant byte transferred first.
The range is 0 to 255 when DATa:WIDth is 1. Center screen is 127. The range is
0 to 65,535 when DATa:WIDth is 2. The upper limit is one division above the
top of the screen and the lower limit is one division below the bottom of the
screen.
 is the same as  except that the byte order is swapped,
meaning that the least significant byte is transferred first. This format is useful
when transferring data to IBM compatible PCs.
 is the same as  except that the byte order is swapped,
meaning that the least significant byte is transferred first. This format is useful
when transferring data to IBM compatible PCs.
Arguments