Command Descriptions
TDS Family Oscilloscope Programmer Manual 2–251
Status and Error
FACtory, *PSC, *RCL, RECAll:SETUp, *SAV, SAVe:SETUp


*RST returns the instrument settings to the factory defaults (see Appendix D:
Factory Initialization Settings).
The *RST command does not alter the following:
HThe state of the IEEE Std 488.1–1987 interface.
HThe selected IEEE Std 488.1–1987 address of the digitizing oscilloscope.
HCalibration data that affect device specifications.
HThe Output Queue.
HThe Service Request Enable Register setting.
HThe Standard Event Status Enable Register setting.
HThe Power-on status clear flag setting.
HAlias definitions.
HStored settings.
HThe *PUD? response.
RS232:BAUd
RS-232/Centronics Hardcopy Interface Only
Sets or queries RS-232-C interface transmission speed.
RS232
 

Group
Related Commands
Syntax
Group