Command Descriptions
2–308 TDS Family Oscilloscope Programmer Manual
Trigger
TRIGger:MAIn:PULse:RUNT?
TRIGger : :PULse
MAIn RUNT ?
:
TRIGGER:MAIN:PULSE:RUNT?
might return :TRIGGER:MAIN:PULSE:RUNT:POLARITY POSITIVE;THRESHĆ
OLD:HIGH 2.00E+0;LOW 800.0E-3.
TRIGger:MAIn:PULse:RUNT:POLarity
Not on TDS 400A
Sets or queries the polarity for the main pulse runt trigger. This command is
equivalent to selecting Polarity in the Trigger menu.
Trigger
TRIGger:MAIn:PULse:RUNT:POLarity { EITher | NEGAtive | POSITIVe }
TRIGger:MAIn:PULse:RUNT:POLarity?
NEGAtive
<Space>
?
EITher
POSITIVe
TRIGger : :PULse
MAIn
RUNT : POLarity
:
NEGAtive indicates that the falling edge crosses the high threshold and the rising
edge recrosses the high threshold without either edge ever crossing the low
threshold.
POSITIVe indicates that the rising edge crosses the low threshold and the falling
edge recrosses the low threshold without either edge ever crossing the high
threshold.
EITher indicates either NEGAtive or POSITIVe polarity.
TRIGGER:MAIN:PULSE:RUNT:POLARITY NEGATIVE
specifies that the polarity of the main pulse runt trigger is negative.
Group
Syntax
Examples
Group
Syntax
Arguments
Examples