Command Descriptions
TDS Family Oscilloscope Programmer Manual 2–175
"!!
!
"!
  
 is from 0 to 100, and is the percent of the waveform that is displayed left
of the center graticule.
In extended-acquisition-length mode, it is the percent of the acquisition that is
displayed left of the center graticule.
 
sets the horizontal position of the waveform such that 10% of the waveform is to
the left of screen center.
HORizontal:RECORDLength
Sets the number of data points that are acquired for each record. This is
equivalent to setting Record Length in the Horizontal menu.
When extended-acquisition-length mode is on, this command can still set and
return values. However, this feature will not actually work while extended-ac-
quisition-length mode is on.
Horizontal
"!! 
"!!
!
  
"!
Table 2–32 shows supported values for <NR1>.
Arguments
Examples
Group
Syntax
Arguments