Command Descriptions
2–226 TDS Family Oscilloscope Programmer Manual
OFF or <NR1> = 0 turns measurements off. You can also turn the state off by
deselecting the source.
ON or <NR1> 0 turns measurements on.
MEASUREMENT:MEAS1:STATE ON
turns measurement defined as MEAS1 on.
MEASUREMENT:MEAS4:STATE?
returns either 0 or 1, indicating the state of MEAS4.
MEASUrement:MEAS<x>:STDdev? (Query Only)
TDS 500C & 700C Only
Returns the standard deviation of values accumulated for this measurement since
the last statistical reset.
Measurement
MEASUrement:MEAS<x>:STDdev?
:MEASUrement MEAS <x> ?
:STDdev
<NR3>
MEASUREMENT:MEAS1:STDDEV?
might return: :MEASU:MEAS1:STD 21.0E-12
MEASUrement:MEAS<x>:TYPe
Sets or queries the measurement type for the measurement specified by
MEAS<x>. This is equivalent to selecting the measurement in the Select
Measurement side menu.
Measurement
MEASUrement:MEAS<x>:TYPe { AMPlitude | AREa | BURst | CARea |
CMEan | CRMs | DELay | EXTINCTRATIO (TDS 500C and 700C) |
EXTINCTPCT (TDS 500C and 700C) | EXTINCTDB (TDS 500C and 700C)
|FALL | FREQuency | HIGH | HITs (TDS 500C and 700C) |LOW |
Arguments
Examples
Group
Syntax
Returns
Examples
Group
Syntax