Command Descriptions
TDS Family Oscilloscope Programmer Manual 2–191
0.0E+0,0.0E+0;:MASK:MASK8:POINTSP 0.0E+0,0.0E+0;:MASK:STAN
OC1;DIS 1;COUN:STATE 0;:MASK:FILT ENA;AUTOS:OFFSETA 1;MOD
MAN;:MASK:SOU CH1;TBP 0.0E+0
MASK:AUTOSet:MODe
TDS 500C & 700C Only
Controls whether or not an autoset will automatically be done after a standard
mask is selected.
Mask
MASK:AUTOSet:MODe { AUTO | MANual }
MASK:AUTOSet:MODe?
<Space>
?
AUTO
MANual
AUTOSet
MASK :MODe
:
AUTO makes autoset run automatically when a standard mask is selected from the
front panel. Autoset does not keep running. It runs once when a standard is
selected.
MANual means that the mask autoset algorithm for a standard mask will be run
only if the user presses the autoset button or sends the AUTOSET START
command. If a standard mask is not currently selected, the mask autoset
algorithm will not run and the normal autoset will run when the user presses the
autoset button. Manual is the default value.
MASK:AUTOSET:MODE AUTO
makes autoset run automatically when a standard mask is selected from the front
panel.
MASK:AUTOSet:OFFSETAdj
TDS 500C & 700C Only
Controls whether or not a standard mask autoset has the freedom to adjust
vertical offset so the waveform matches the mask. This can be needed if ground
(0 V) in the oscilloscope does not match ground for the user’s signal. For the
Group
Syntax
Arguments
Examples