Command Descriptions
TDS Family Oscilloscope Programmer Manual 2–107
*RCL, RECAll:SETUp, *RST, *SAV, SAVe:SETUp, TEKSecure
DELEte:SETUp { <NR1> | ALL }
DELEte : SETUp <Space>
ALL
<NR1>
<NR1> is a value in the range 1 to 10, and specifies a setup storage location.
Using an out-of-range value causes an execution error.
ALL specifies all the stored setups.
DELETE:SETUP ALL
removes all stored setups. All ten storage locations are initialized to the factory
default setup.
DELEte:WAVEform (No Query Form)
Deletes one or all of the stored reference waveforms from memory. The memory
allocated for the reference location is then available for reallocation. Memory
must be reallocated for the deleted references before any waveform data can be
stored in the reference location.
NOTE. The waveform data is not actually cleared from the reference location.
Save and Recall
RECAll:WAVEform, SAVe:WAVEform, TEKSecure
DELEte:WAVEform { REF<x> | ALL }
DELEte : <Space>
ALL
<x>
WAVEform
REF
REF<x> (REF1, REF2, REF3 or REF4) specifies one of the reference memory
locations.
ALL specifies all the stored waveforms.
Related Commands
Syntax
Arguments
Examples
Group
Related Commands
Syntax
Arguments