Command Descriptions
2–310 TDS Family Oscilloscope Programmer Manual
TRIGger:MAIn:PULse:RUNT:THReshold:HIGH
TDS 510A, 500C, 600B, & 700C Only
Sets or queries the upper limit for the main pulse runt trigger. This command is
equivalent to setting the threshold in the Pulse Runt Threshold side menu.
Trigger
TRIGger:MAIn:PULse:RUNT:THReshold:HIGH <NR3>
TRIGger:MAIn:PULse:RUNT:THReshold:HIGH?
TRIGger ::PULse
MAIn RUNT :
:
?
<Space> <NR3>
HIGH
THReshold
<NR3> is the threshold, in volts.
TRIGGER:MAIN:PULSE:RUNT:THRESHOLD:HIGH 120E-3
sets the upper limit of the pulse runt trigger to 120 mV.
TRIGger:MAIn:PULse:RUNT:THReshold:LOW
TDS 510A, 500C, 600B, & 700C Only
Sets or queries the lower limit for the main pulse runt trigger. This command is
equivalent to setting the threshold in the Pulse Runt Threshold side menu.
Trigger
TRIGger:MAIn:PULse:RUNT:THReshold:LOW <NR3>
TRIGger:MAIn:PULse:RUNT:THReshold:LOW?
Group
Syntax
Arguments
Examples
Group
Syntax