Command Descriptions
2–252 TDS Family Oscilloscope Programmer Manual
BAUd
:
<Space>
?
RS232
<NR1>
<NR1> where <NR1> can be 300, 600, 1200, 2400, 4800, 9600 or 19200.
RS232:BAUD 9600
sets the transmission rate to 9600 baud.
RS232:HARDFlagging

RS-232/Centronics Hardcopy Interface Only

Sets or queries the input and output hard flagging over the RS-232 port. It uses
the RFR (Ready For Receive) and CTS (Clear To Send) lines to control data
transmission. On output, the oscilloscope transmits data only when CTS is
asserted. When CTS is not asserted, the oscilloscope stops transmitting data. On
input, it asserts RFR until the receive queue is full. Then it unasserts RFR to stop
transmission from an external printer. CTS remains unasserted until the receive
queue is not full. At that time, CTS is asserted again to restart transmission.
RS232
RS232:HARDFlagging { ON | OFF | <NR1> }
RS232:HARDFlagging?
:
RS232 HARDFlagging
<Space>
ON
?
OFF
<NR1>
<ON> or <NR1>0 turn on hardflagging.
<OFF> or <NR1> = 0 turn off hardflagging.
RS232:HARDFLAGGING ON
turns on hard flagging.
Arguments
Examples
Group
Syntax
Arguments
Examples