Command Descriptions
2–136 TDS Family Oscilloscope Programmer Manual
FACtory (No Query Form)
Resets the digitizing oscilloscope to its factory default settings. This command is
equivalent to selecting Recall Factory Setup in the Waveform Save/Recall
menu.
Miscellaneous
*PSC, *RCL, RECAll:SETUp, *RST, *SAV, SAVe:SETUp


Setting the digitizing oscilloscope to factory default includes:
HClears the Event Status Enable Register.
HClears the Service Request Enable Register.
HSets the Device Event Status Enable Register to 255.
HSets the Power On Status Clear Flag to TRUE
HPurges all defined aliases.
HEnables all Command Headers (HEADer ON).
HSets the macro defined by *DDT to a “zero-length field.”
HClears the pending operation flag and associated operations.
The FACtory command does not alter the following:
HThe state of the GPIB (IEEE Std 488.1-1987) interface.
HThe selected GPIB address.
HCalibration data that affects device specifications.
HProtected user data.
HStored settings.
HThe current password (if implemented).
Group
Related Commands
Syntax