Troubleshooting

 

 

10Troubleshooting

10.1 System error messages

Code

Description

Possible cause

Remedy

1

Common line open

Ground (common) line has open

Check connector

 

 

circuit. All temperature element

attachment on the module;

 

 

signals will be disabled or

then check continuity on

 

 

inaccurate.

common (black & white)

 

 

 

cable

3~39

Element open

Temperature element signal

Check connector

 

 

cable

attachment on the module;

 

 

( #1 ~ 6) has open circuit.

then check continuity on

 

 

 

pointed signal cable (#1 ~

 

 

 

6)

4~40

Element short

Temperature element signal

Disengage connector from

 

 

cable

the module; then check

 

 

( #1 ~ 6) has short circuit.

continuity on pointed

 

 

 

signal cable (#1 ~ 6)

23

#0 element over

When reference #0 element has

Check power supply

 

range

more than ±1.1°C deviation

voltage on 4532 ATC HART

 

 

from 0°C.

terminal H+ and H-

24

Memory defect

When a defect was discovered

Replace main CPU board

 

(ROM)

during whole memory

 

 

 

parameter check. Cyclic data

 

 

 

comparison between previous

 

 

 

check sum to current one.

 

29

Element exposed

Liquid level dropped below #1

No liquid temperature

 

 

element position.

measurement is available.

32

Low power supply

Supply voltage on multi drop

Check power supply on the

 

 

HART loop to 4532 ATC is below

host instrument and

 

 

16VDC from designated host

consumption of connected

 

 

instrument.

loop powered HART device

41

Memory defect

Fault during Write and Read

Replace main CPU board

 

(RAM)

sequence, not completed.

 

42

Memory defect

Fault during Write and Read

Check the write command

 

(EEROM)

sequence, not completed.

itself that is acceptable to

 

 

 

4532 ATC; if command is

 

 

 

OK, replace main CPU

 

 

 

board

These error codes are displayed on the ToF tool display when the ToF tool is properly connected. For details regarding the error display on host instrument, please refer to documentation of 4590 TSM or 6000 Series Servo Tank Gauge.