VT1422A Command Reference 269Chapter 7
CALibration
Related Commands: SYST:ERR?, CAL:SETup, *CAL?
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Usage See CAL:SETup.

CALibration:STORe

CALibration:STORe <type> stores the VT1422A’s most recently measured
calibration constants into Flash Memory (Electrically Erasable Programmable Read
Only Memory). When <type>=ADC, the module stores its A/D calibration constants
as well as constants generated from *CAL?/CAL:SETup into Flash Memory.
When <type>=TARE, the module stores the most recently measured CAL:TARE
channel offsets into Flash Memory.
Note The VT1422A's Flash Memory has a finite lifetime of approximately 10,000 write
cycles (unlimited read cycles). While executing CAL:STOR once every day would
not exceed the lifetime of the Flash Memory for approximately 27 years, an
application that stored constants many times each day would unnecessarily shorten
the Flash Memory's lifetime. See Comments below.
Parameters

Comments The Flash Memory Protect jumper (JM2201) must be set to the enable position

before executing this command (See “Disabling Flash Memory Access
(Optional)” on page 29).
Channel offsets are compensated by the CAL:TARE command even when not
stored in the Flash Memory. There is no need to use the CAL:STORE TARE
command for channels which are re-calibrated frequently.
When Accepted: Not while INITiated
Related Commands: CAL:VAL:RES, CAL:VAL:VOLT
*RST Condition: Stored calibration constants are unchanged
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage CAL:STORE ADC Store cal constants in non-volatile
memory after A/D calibration
CAL:STORE TARE Store channel offsets in non-volatile
memory after channel tare
Parameter
Name Parameter
Type Range of
Values Default
Units
type discrete (string) ADC | TARE none