VT1422A Command Reference 297Chapter 7
INPut
INPut
The INPut subsystem controls configuration of programmable input Signal
Conditioning Plug-Ons (SCPs).
Subsystem Syntax INPut
:FILTer
[:LPASs]
:FREQuency <cutoff_freq>,(@<ch_list>)
:FREQuency? (@<channel>)
[:STATe] 1 | 0 | ON | OFF,(@<channel>)
[:STATe]? (@<channel>)
:GAIN <chan_gain>,(@<ch_list>)
:GAIN? (@<channel>)
:LOW <wvolt_type>,(@<ch_list>)
:LOW? (@<channel>)
:POLarity NORMal | INVerted,(@<ch_list>)
:POLarity? (@<channel>)

INPut:FILTer[:LPASs]:FREQuency

INPut:FILTer[:LPASs]:FREQuency <cutoff_f req>,(@<ch_list>) sets the cutoff
frequency of the filter on the specified channels.
Parameters
Comments The <cutoff_freq> parameter may be specified in kilohertz (kHz). A
programmable Filter in either an SCP or a Remote Signal Conditioning unit
(RSC) has a choice of several discrete cutoff frequencies. The cutoff frequency
set will be the one closest to the value specified by <cutoff_freq>.
For the VT1529A/B, this command affects an 8-channel bank, not a single
channel.
Sending MAX for the <cutoff_freq> selects the SCP or RSC's highest cutoff
frequency. Sending MIN for the <cutoff_freq> selects the SCP or RSC's lowest
cutoff frequency. To disable filtering (the "pass through" mode), execute the
INP:FILT:STATE OFF command.
Sending a value greater than the SCP's highest cutoff frequency or less than the
SCP's lowest cutoff frequency generates a -222 "Data out of range" error.
Parameter
Name Parameter
Type Range of
Values Default
Units
cutoff_freq numeric (float32)
(string) see comment |
MIN | MAX Hz
ch_list channel list (string) 100 - 15731 none