VT1422A Command Reference 401Chapter 7
IEEE-488.2 Common Command Reference
*ESE
*ESE <mask> Standard Event Status Enable Register Command. Enables one or
more events in the Standard Event Status Register to be reported in bit 5 (the Standard
Event Status Summary Bit) of the Status Byte Register. An event is enabled by
specifying its decimal weight for <mask>. To enable more than one event (bit),
specify the sum of the decimal weights. The data type for <mask> is int16.
Send with VXIplug&play Function: hpe1422_cmd(...)
*ESE?
*ESE? Standard Event Status Enable Query. Returns the weighted sum of all enabled
(unmasked) bits in the Standard Event Status Register. The data type for this returned
value is int16.
*ESR?
*ESR? Standard Event Status Register Query. Returns the weighted sum of all set
bits in the Standard Event Status Register. After reading the register, *ESR? clears
the register. The events recorded in the Standard Event Status Register are
independent of whether or not those events are enabled with the *ESE command to
set the Standard Event Summary Bit in the Status Byte Register. The Standard Event
bits are described in the *ESE command. The data type for this returned value is int16.
*GMC?
*GMC? <name> Get Macro query. Returns arbitrary block response data which
contains the command or command sequence defined for <name>. For more
information, see “Arbitrary Block Program and Response Data” on page233.
Bit # 7 6 5 4 3 2 1 0
Weighted Value 128 64 32 16 8 4 2 1
Event power-On User
Request Command
Error Execution
Error Device Dependent
Error Query
Error Request
Control Operation
Complete