258 VT1422A Command Reference Chapter 7
ARM
While ARM:SOUR is IMM, simply INITiate the trigger system to start a
measurement scan.
When Accepted: Before INIT only.
Related Commands: ARM:IMM, ARM:SOURCE?, INIT[:IMM],
TRIG:SOUR
*RST Condition: ARM:SOUR IMM
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage ARM:SOUR BUS Arm with ARM command
ARM:SOUR TTLTRG3 Arm with VXIbus TTLTRG3 line
ARM:SOURce?
ARM:SOURce? returns the current arm source configuration. See the ARM:SOUR
command for more response data information.
Returned Value: Discrete, one of BUS, HOLD, IMM, SCP, or TTLT0 through
TTLT7. The data type is string.
Send with VXIplug&play Function: hpe1422_cmdString_Q(...)
Usage ARM:SOUR? An enter statement return arm
source configuration