VT1422A Command Reference 345Chapter 7
[SENSe]

[SENSe:]FUNCtion:HVOLtage

[SENSe:]FUNCtion:HVOLtage [<range>,](@<ch_list>) links the specified
VT1529B channels to return dc voltage on the high-level input (Excitation Sense +
and -) pins on the RJ-45 connector for the channel.
This command is valid only on the VT1529B and is available with VXIplug&play
driver revision A.01.09 or later (revision A.01.06 or later of hpe1422_32.dll).
Parameters

Comments The VT1422A has five ranges: 0.0625V dc, 0.25 Vdc, 1 Vdc, 4 Vdc, and

16 Vdc. To select a range, simply specify the range value (for example, 4 selects
the 4 Vdc range). If a value is specified that is larger than one of the first four
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the
16 Vdc range). Specifying a value larger than 16 causes an error. Specifying 0
selects the lowest range (0.0625 Vdc). Specifying AUTO selects auto range.
The default range (no range parameter specified) is auto range.
The VT1529A/B has a fixed gain of 32. Keep this in mind when <range> is set.
See “DCV Measurement Command Sequence” on page 177.
When Accepted: Not while INITiated
Related Commands: CAL:REMote?
*RST Condition: SENSE:FUNC:VOLT AUTO on all channels and bridges set
to FBEN on all strain channels.
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage FUNC:HVOL (@10000:10001) Channels 0 - 1 measure high-level input
voltage in auto-range (defaulted)
Parameter
Name Parameter
Type Range of
Values Default
Units
range numeric (float32) see comments Vdc
ch_list channel list(string) 10000 - 15731 none