VT1422A Command Reference 287Chapter 7
DIAGnostic
Returned Value: DIAG:REM:USER:DATA? returns an IEEE definite length
data block which represents an array of 894, int16 values.
RST Condition: Stored values not changed by *RST
Send with VXIplug&play Function: hpe1422_cmdInt16Arr_Q(...)

DIAGnostic:TEST:REMote:NUMber?

DIAGnostic:TEST:REMote:NUMber? <test_num>,<iterations>,(@<channel>)
executes a selected self-test number on a single Remote Signal Conditioning Unit
connected through the VT1539A SCP. See DIAG:TEST:REM:SELF? for details of
each test.
Parameters

Comments The <test_num> parameter specifies the test to perform. For explanations of the

test numbers, See “DIAGnostic:TEST:REMote:SELFtest?” on page 288.
The <iterations> parameter specifies the number of times to perform a test.
The <channel> parameter may contain only any single channel number on a
Remote Signal Conditioning Unit. All channels on that RSCU will be tested.
Returned Value:
The data type for this returned value is int16.
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Parameter
Name Parameter
Type Range of
Values Default
Units
test_num numeric 1 - 5 none
iterations numeric 1 - 32767 none
channel channel list (string) 10000 - 15731 none
Value Meaning Further Action
0 test passed None
1 number of times test
failed Error information in FIFO buffer.
See error codes below.
-1 Couldn’t start
remote self-test Query the Error Queue (SYST:ERR?)
See error messages starting on page 453.