316 VT1422A Command Reference Chapter 7
OUTPut

OUTPut:POLarity?

OUTPut:POLarity? (@<channel>) returns the polarity on the digital output channel
in <channel>.
Parameters
Comments The <channel> parameter must specify a single channel.
Returned Value: returns one of NORM or INV. The type is string.
Send with VXIplug&play Function: hpe1422_cmdString_Q(...)

OUTPut:SHUNt

OUTPut:SHUNt <enable>,(@<ch_list>) adds shunt resistance to one leg of bridge
on Strain Bridge Completion SCPs and the VT1529A/B Remote Strain Bridge unit.
This can be used for diagnostic purposes and characterization of bridge response.
Parameters

Comments If <ch_list> specifies a non strain SCP, a 3007 "Invalid signal conditioning

plug-on" error is generated.
Only one channel on any one VT1529A/B can be specified in <ch_list>. This
is because a single resistor is used to shunt each of a VT1529A/Bs 32 channels.
The <ch_list> may specify one channel on each of several VT1529A/Bs.
When Accepted: Not while INITiated
Related Commands: [SENSe:]FUNCtion:STRain, [SENSe:]STRain
*RST Condition: OUTP:SHUNT 0 on all Strain SCP and RSC channels
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage OUTP:SHUNT 1,(@116:119) add shunt resistance at channels 16
through 19
Parameter
Name Parameter
Type Range of
Values Default
Units
channel string 100 - 163 none
Parameter
Name Parameter
Type Range of
Values Default
Units
enable boolean (uint16) 0 | 1 | ON | OFF none
ch_list channel list(string) 100 - 15731 none