VT1422A Command Reference 285Chapter 7
DIAGnostic

DIAGnostic:OTDetect[:STATe]?

DIAGnostic:OTDetect[:STATe]? (@<channel>) returns the current state of
"Open Transducer Detection" for the SCP containing the specified <channel>.
Parameters
Comments The <channel> parameter must specify a single channel only.
Returned Value: Returns 1 (enabled) or 0 (disabled). The data type is int16.
Related Commands: DIAG:OTDETECT:STATE ON | OFF
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage DIAG:OTD:STATE? (@108) enter statement returns either a 1 or a 0

DIAGnostic:QUERy:SCPREAD?

DIAGnostic:QUERy:SCPREAD? <reg_addr> returns data word from a Signal
Conditioning Plug-on register.
Parameters

Comments NOTE: This command may not be used while instrument is INITed.

Returned Value: returns numeric register value. data type is int32.
Send with VXIplug&play Function: hpe1422_cmdInt32_Q(...)
Usage DIAG:QUERY:SCPREAD? 258 read SCP #4 config/status register
enter statement here return SCP ID value
Parameter
Name Parameter
Type Range of
Values Default
Units
channel channel list (string) 100 - 163 none
Parameter
Name Parameter
Type Range of
Values Default
Units
reg_addr numeric (int32) 0-65,535 none