VT1422A Command Reference 263Chapter 7
CALibration
Figure 7-3. Levels of Working Calibration
Subsystem Syntax CALibration
:CONFigure
:RESistance
:VOLTage <range >, ZERO | FS
:REMote?
:DATA <cal_data_block>
:DATA?
:STORe
:SETup
:SETup?
:STORe ADC | TARE
:TARE (@<ch_list>)
:RESet
:TARE?
:VALue
:RESistance <ref_ohms>
:VOLTage <ref_volts>
:ZERO?
CALibration:CONFigure:RESistance
CALibration:CONFigure:RESistance connects the on-board reference resistor to
the Calibration Bus. A four-wire measurement of the resistor can be made with an
external multimeter connected to the HCAL, LCAL, HOHM and LOHM terminals
on the Terminal Module or the VH, VL, ΩH, and ΩL terminals on the Cal Bus
connector when not using a terminal module.
SCP
SCP
SCP
SCP
SCP
8
8
8
8
8
8
8
8
SCP
SCP
SCP
Multiplexe r
CAL:ZERO?
CAL:TARE
*CAL? or
CAL:SETup
User’s System Wiring