302 VT1422A Command Reference Chapter 7
INPut

INPut:LOW?

INPut:LOW? (@<channel>) returns the LO input configuration for the channel
specified by <channel>. This command is for strain SCPs only, not for VT1529A/B.
Parameters
Comments The <channel> parameter must specify a single channel only.
Returned Value: Returns FLO or WV. The data type is string.
Related Commands: INP:LOW
Send with VXIplug&play Function: hpe1422_cmdString_Q(...)
Usage INP:LOW? (@103) enter statement will return either FLO or
WV for channel 3

INPut:POLarity

INPut:POLarity <mode>,<ch_list> sets logical input polarity on a digital
SCP channel.
Parameters

Comments If the channels specified are on an SCP that doesn't support this function, an

error will be generated. See the SCP's User's Manual to determine its capabilities.
Related Commands: for output sense; SOURce:PULSe:POLarity
*RST Condition: INP:POL NORM for all digital SCP channels.
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage INP:POL INV,(@140:143) invert first 4 channels on SCP at SCP
position 5. Channels 40 through 43
Parameter
Name Parameter
Type Range of
Values Default
Units
channel channel list (string) 100 - 163 none
Parameter
Name Parameter
Type Range of
Values Default
Units
mode discrete (string) NORMal | INVerted none
ch_list string 100 - 163 none