VT1422A Command Reference 277Chapter 7
DIAGnostic

DIAGnostic:CALibration:SETup[:MODE]

DIAGnostic:CALibration:SETup[:MODE] <mode> sets the type of calibration to
use for analog output SCPs like the VT1531A and VT1532A when *CAL? or
CAL:SET are executed.
Parameters

Comments When <mode> is set to 1 (the *RST Default) channels are calibrated using the

Least Squares Fit method to provide the minimum error overall (over the entire
output range). When <mode> is 0, channels are calibrated to provide the
minimum error at their zero point. See the SCPs User's Manual for its accuracy
specifications using each mode.
Related Commands: *CAL?, CAL:SET, DIAG:CAL:SET:MODE?
*RST Condition: DIAG:CAL:SET:MODE 1
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage set analog DAC SCP cal mode for best zero accuracy
DIAG:CAL:SET:MODE 0 set mode for best zero cal
*CAL? start channel calibration

DIAGnostic:CALibration:SETup[:MODE]?

DIAGnostic:CALibration:SETup[:MODE]? returns the currently set calibration
mode for analog output DAC SCPs.

Comments Returns a 1 when channels are calibrated using the Least Squares Fit method to

provide the minimum error overall (over the entire output range). Returns a 0
when channels are calibrated to provide the minimum error at their zero point.
See the SCPs User's Manual for its accuracy specifications using each mode.
The data type is int16.
Related Commands: DIAG:CAL:SET:MOD, *CAL?, CAL:SET
*RST Condition: DIAG:CAL:SET:MODE 1
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Parameter
Name Parameter
Type Range of
Values Default
Units
mode boolean (uint 16) 0 | 1 volts