VT1422A Command Reference 385Chapter 7
STATus

STATus:OPERation:PTRansition

STATus:OPERation:PTRansition <transition_mask> sets bits in the Positive
Transition Filter (PTF) register. When a bit in the PTF register is set to one, the
corresponding bit in the Condition register must change from a zero to a one in order
to set the corresponding bit in the Event register. When a bit in the PTF register is
zero, a positive transition of the Condition register bit will not change the Event
register bit.
Parameters
Comments <transition_mask> may be sent as decimal, hex (#H), octal (#Q), or binary (#B).
If both the STAT:OPER:PTR and STAT:OPER:NTR registers have a
corresponding bit set to one, any transition, positive or negative, will set the
corresponding bit in the Event register.
If neither the STAT:OPER:PTR or STAT:OPER:NTR registers have a
corresponding bit set to one, transitions from the Condition register will have
no effect on the Event register.
Related Commands: STAT:OPER:PTR?, STAT:OPER:NTR
Set to all ones by: STAT:PRESet and power-on.
*RST Condition: No change
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage STAT:OPER:PTR 16 When "Measuring" bit goes true, set bit 4
in Status Operation Event register.

STATus:OPERation:PTRansition?

STATus:OPERation:PTRansition? returns the value of bits set in the Positive
Transition Filter (PTF) register.
Comments Returned Value: Decimal weighted sum of all set bits. The data type is uint16.
Related Commands: STAT:OPER:PTR
*RST Condition: No change
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
Usage STAT:OPER:PTR? Enter statement returns current value of
bits set in the PTF register
Parameter
Name Parameter
Type Range of
Values Default
Units
transition_mask numeric (uint16) 0 - 32767 none