Sunrise Medical 515 SERIES manual Minor Troubleshooting Chart, Symptom Possible Cause Remedy

Page 19

English

19

MINOR TROUBLESHOOTING

The following troubleshooting chart will help you analyze and correct minor oxygen concentrator malfunctions. If the suggested procedures do not help, switch to your reserve oxygen system and call your Sunrise Medical homecare provider. Do not attempt any other maintenance.

WARNING–To avoid electric shock hazard, do not remove the cabinet. The cabinet should only be removed by a qualified Sunrise Medical homecare technician.

MINOR TROUBLESHOOTING CHART

Symptom

 

Possible Cause

 

Remedy

 

 

 

 

A. Unit does not operate. Power light is off when the power

1.

Power cord not properly inserted into wall outlet.

1.

Check power cord connection at the wall outlet. On 230 volt units,

 

switch is “On.” Audible alarm is pulsing and Service Required

 

 

 

also check the back of the unit.

 

light is flashing.

2.

No power at wall outlet.

2.

Check your home circuit breaker and reset if necessary. Use a

 

 

 

 

 

different wall outlet if the situation occurs again.

 

 

3.

Oxygen concentrator circuit breaker activated (selected units).

3.

Press the concentrator circuit breaker reset button (if equipped)

 

 

 

 

 

located below the power switch. Use a different wall outlet if the

 

 

 

 

 

situation occurs again.

 

 

 

 

 

If the above remedies do not work, contact your

 

 

 

 

 

Sunrise Medical provider.

 

 

 

 

 

 

 

 

B. Unit operates, the Power light is on

1.

Air filter is blocked.

1.

Check the air filter. If the filter is dirty, wash it following the

 

when the power switch is “On.” Red Service Required light is

 

 

 

cleaning instructions on page 18.

 

illuminated. Audible alarm may be sounding.

2.

Exhaust is blocked.

2.

Check the exhaust area; make sure there is nothing restricting

 

 

 

 

 

the unit exhaust.

 

 

3. Blocked or defective cannula, catheter, face mask,

3.

Detach cannula, catheter, or face mask. If proper flow is restored,

 

 

 

or oxygen tubing.

 

clean or replace if necessary. Disconnect the oxygen tubing at the

 

 

 

 

 

oxygen outlet. If proper flow is restored, check oxygen tubing for

 

 

 

 

 

obstructions or kinks. Replace if necessary.

 

 

4.

Blocked or defective humidifier bottle.

4.

Detach the humidifier from the oxygen outlet. If proper flow is

 

 

 

 

 

obtained, clean or replace humidifier.

 

 

5.

Flow meter set too low.

5.

Set flow meter to prescribed flow rate.

 

 

 

 

 

If the above remedies do not work, contact your

 

 

 

 

 

Sunrise Medical provider.

 

 

 

 

 

 

 

C. If any other problems occur with your oxygen concentrator.

 

 

1.

Turn your unit “Off.” Switch to your reserve oxygen system, and

 

 

 

 

 

contact your Sunrise Medical provider immediately.

 

 

 

 

 

 

 

 

MINOR TROUBLESHOOTING CHART - ONLY OSD Concentrators

 

Visual Indicator Lights

 

Possible Cause

 

Remedy

 

 

 

 

A. Both the green Normal Oxygen and the yellow Low Oxygen lights are

1.

OSD malfunction.

1.

Contact your Sunrise Medical provider.

 

 

 

either on or off.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

B. Yellow Low Oxygen light is on or the yellow Low Oxygen light is on

1.

Flow meter is not properly set.

1.

Ensure the flow meter is properly set to the prescribed number.

 

 

 

and the intermittent audible signal is sounding.

2.

Air filter is blocked.

2.

Check the air filter. If the filter is dirty, wash it following the

 

 

 

 

 

 

 

cleaning instructions on page 18.

 

 

 

 

3.

Exhaust is blocked.

3.

Check the exhaust area; make sure there is nothing restricting the

 

 

 

 

 

 

 

unit exhaust.

 

 

 

 

 

 

 

 

 

 

 

 

 

C. Red Service Required light is on and an intermittent audible sig-

1.

Flow meter is not properly set.

1.

Ensure the flow meter is properly set to the prescribed number.

 

 

 

nal is sounding.

2.

Air filter is blocked.

2.

Check the air filter. If the filter is dirty, wash it following the

 

 

 

 

 

 

 

cleaning instructions on page 18.

 

 

 

 

3.

Exhaust is blocked.

3.

Check the exhaust area; make sure there is nothing restricting the

 

 

 

 

 

 

 

unit exhaust.

 

 

 

 

 

 

 

If the above remedies do not work, contact your

 

 

 

 

 

 

 

Sunrise Medical provider.

 

 

 

 

 

 

 

 

 

 

 

 

SP-515

Image 19
Contents Guide d’instructions Concentrateur ’oxygène DeVilbiss Table DES Matières Table of ContentsInhaltsverzeichnis Catalog Number 515UK 515DS, 515DZ 515KS, 515KZ, 515NSClass II Equipment Double Insulated Type B Applied Part Serie DE 5 Litros DE Devilbiss EspecificacionesEquipement de classe II à isolation double CaractéristiquesEN 60601+A1+A2 Klasse II, Ausrüstung doppelt isoliert Numero di catalogo SpecificheNessuna diminuzione di prestazione Konformitätserklärung Declaration of ConformityDeclaración DE Conformidad Déclaration DeconformitéNOTA-Informazione che merita partico- lare attenzione Important SafeguardsFunktionsweise Ihres DeVilbiss-O2Konzentrators Por qué su médico le prescribió un suplemento de oxígenoCome funziona il Concentratore di ossigeno DeVilbiss Front View Vista frontal ConcentrateurPartie avant Vorderansicht Vista dal lato anteriore Back View Vista posterior AccessoriesAccessoires AccessoriDEN Konzentrator AUF- Stellen Installazione DEL ConcentratoreAntes DE Usar SU Concentrador Avant D’UTILISER Votre CON- CentrateurÜber Conexión tubo de oxígeno con humidificaciónOxygen Tubing Connection With Humidification Connexion du tube d’oxy- gène à l’humidificateurNUR DEVILBISS- O2 KONZENTRA- Toren MIT OSD Signal will briefly alarmUSO DE SU Concentrador Utilisation DE Votre ConcentrateurÜberprüfen Sie den Oxygen Concentrator ReservaSigeno Entretien DE VotreCánula, Mascarilla y Tubos Air Filter and Oxygen Outlet ConnectorCannula, Mask, and Tubing Exterior CabinetMinor Troubleshooting Chart Minor TroubleshootingSymptom Possible Cause Remedy Su proveedor Sunrise Medical Detección Y Solución DE Problemas MenoresTabla DE Detección Y Solución DE Problemas Menores Síntoma Causa posible SoluciónDépannage Mineur Symptômes Cause Possible SolutionTableau DE Dépannage Mineur Wenn diese Abhilfemaßnahmen nichts bewirken Fehlersuche BEI Kleineren ProblemenDiagramm ZUR Fehlersuche BEI Kleineren Problemen Symptom Mögliche Ursache AbhilfeSintomo Possibile Causa Rimedio Individuazione DI Problemi MinoriTabella DI Individuazione DI Problemi Minori Se i rimedi indicati non risolvono il problemaWortlichen Ihren zuständigen DeVilbiss Fachhändler SP-515 SP-515 Page 44-138-444-6688