SECTION 3—OPERATION

Activating a Mechanical Emergency Release

Primary Emergency Release

NOTE: For this procedure, refer to FIGURE 3.2.

NOTE: This procedure will bring the boom down or lift it up if the hand control is not functioning properly.

To activate the primary emergency release, insert a pen into the hole labeled Emergency Up or Emergency Down on the control box.

Secondary Emergency Release

NOTE: For this procedure, refer to FIGURE 3.3.

NOTE: All patient lift actuators are equipped with a mechanical emergency release. The mechanical release will enable the actuator to retract without power.

NOTE: Use the primary emergency release first before using the secondary emergency release procedure. This procedure should only be used if the primary emergency release procedure is not functioning or is unreachable.

NOTE: The lift MUST be under a load for the mechanical release to function.

To activate the secondary emergency release, pull up on the RED emergency grip and pull down on the boom at the same time.

Charging the Battery

NOTE: For this procedure, refer to FIGURE 3.4.

NOTE: Invacare recommends the battery be recharged daily to prolong battery life.

NOTE: An audible alarm will sound when battery is low.

1.Lift up on the handle on the back of the battery.

2.Lift the battery up and away from the control box.

Control Box

Emergency Up

Emergency Down

Hole

Hole

FIGURE 3.2 Primary Emergency Release

RED

Emergency Grip

CAUTION

Make sure there is an audible click when mount- ing battery on the battery charger to ensure proper mounting. Otherwise, injury or damage may occur.

3.Place the battery on the battery charger. Push the top of the battery against the mounting bracket until there is an audible click.

NOTE: The charge LED will illuminate. When charged, the LED will stop illuminating.

NOTE: It will take approximately four hours to charge a battery that requires a full charge.

4.Lift up on the handle on the back of the battery.

5.Lift the battery up and away from the battery charger.

CAUTION

Make sure there is an audible click when mount- ing battery onto the control box to ensure proper mounting. Otherwise, injury or damage may occur.

6.Place the battery on the control box. Push the top of the battery against the mounting bracket until there is an audible click.

FIGURE 3.3 Secondary Emergency Release

Battery

Handle

(STEPS 1, 2, 4 and 5)

Battery Charger

Control Box

(STEP 6)

(STEP 3)

 

An audible “click” will be heard when properly installed (STEPS 3 and 6)

FIGURE 3.4 Charging the Battery

Part No 1150704

13

Jasmine™ Patient Lift

Image 13
Contents Jasmine Patient Lift Symbol Legend Accessories WarningDate of Manufacture Table of Contents Accessories Troubleshooting MaintenanceLimited Warranty Transferring the PatientSpecial Notes Maintenance Must be performed only by qualified personnelSling Sizes Label LocationJasmine Patient Lift Product ParametersJasmine Scale JLS5 Using the Sling Weight LimitationGeneral Guidelines Assembling the LiftLifting the Patient Performing MaintenanceTransferring the Patient Assembling the Mast Assembly to the Base AssemblyAssembling the Patient Lift Attaching the Battery Charger to Mast Assembly Attaching the Battery Charger Mounting Bracket to the WallIng battery on the battery charger to confirm May occurOperating the Patient Lift OperationActivating a Mechanical Emergency Release Charging the BatteryPreparing the Lift for Use Lifting the PatientPositioning the Lift for Use Attaching a Sling to the Swivel Bar Lift the patient. Refer to Lifting/Moving the Patient Lifting/Moving the PatientLifting/Moving the Patient Transferring to a Commode Chair Transferring the PatientTransferring to a Wheelchair Transferring to a Standard CommodeSymptoms Faults Solution TroubleshootingMaintenance Safety Inspection Checklist MaintenanceLubricating the Lift Cleaning the Sling and the Lift Replacing the Boom ActuatorDetecting Wear and Damage Replacing the Boom Actuator Patient Lift Replacing the Leg ActuatorsReplacing the Swivel Bar or Optional Cradle Attachment Replacing the Swivel Bar or Optional Cradle AttachmentChecking and Tightening Mast Pivot Checking and Tightening Mast PivotRemoving the Swivel Bar/Cradle Attachment AccessoriesInstalling the Jasmine Scale Installing the Swivel Bar or Cradle AttachmentOperating the Scale KEY Indicator Displayed Indicator Location DefinitionWeighing the Patient Keypad FunctionsReplacing the Battery Replacing the BatterySymptom Probable Cause Solution TroubleshootingCalibrating the Jasmine Scale Display CodesAccessories Accessories Limited Warranty Lève-personne Jasmine Légende DE Symboles Mise EN Garde Sappliquant AUX AccessoiresMise EN Garde Directives Générales Table DES MatièresLE Montage FonctionnementDépannage Entretien AccessoiresGarantie Limitée Seule une personne qualifiée Doit exécuter lentretien Remarques SpécialesAvertissement Emplacement DES ÉtiquettesLève-personne Jasmine Paramètres DU ProduitPèse-personne Jasmine JLS5 Assemblage du lève-personne Directives GénéralesPoids maximum Lutilisation de la toileLe soulèvement du patient Lutilisation du lève-personneLe transfert du patient LentretienLe montage du lève-personne LE MontageLe montage du mât à la base La fixation du chargeur de pile au montage du mât La fixation du support de chargeur de pile au murLe fonctionnement du lève-personne FonctionnementLe soulèvement/labaissement du lève-personne Lactivation du déverrouillage mécanique durgence Le chargement de la batterieLa préparation du lève-personne aux fins dutilisation LE Soulèvement DU PatientLe positionnement du lève-personne aux fins dutilisation La fixation dune toile à la barre pivotanteLa fixation dune toile à la barre pivotante La fixation dune toile à lattachement facultatif du berceauLa fixation dune toile à lattachement facultatif du berceau Le soulèvement/déplacement du patientLe soulèvement/déplacement du patient Le transfert vers une chaise daisance LE Transfert DUN PatientLe transfert vers un fauteuil roulant Le transfert vers une chaise daisance standardDépannage Symptômes Défauts SolutionListe de contrôle de l’entretien EntretienArticle InstitutionnelLa lubrification du lève-personne Le remplacement de lactionneur de la ceintureDétection de lusure et de lendommagement Nettoyage de la toile et du lève-personneLe remplacement de lactionneur de la ceinture VU DU Dessous Le remplacement des actionneurs de jambesRetirez les capuchons latéraux Vérification et serrage du pivot du mât Vérification et serrage du pivot du mâtRetrait de la barre pivotante/lattache du berceau AccessoiresDétail B Attacher LA Barre DE Pivot Linstallation du pèse-personne JasmineLemploi du pèse-personne Affichage DE Emplacement DE Touche Lindicateur DéfinitionLes fonctions du clavier La pesée du patient Le calibrage du pèse-personne Jasmine Le remplacement de la batterieDépannage Symptômes Cause Probable SolutionLes codes daffichage Remarques USA Garantie Limitée