Invacare 1150704 Troubleshooting, Calibrating the Jasmine Scale, Symptom Probable Cause Solution

Page 31

SECTION 8—ACCESSORIES

Calibrating the Jasmine Scale

NOTE: The Jasmine Scale will be pre‐calibrated at the factory with the load cell. Should it be necessary to re‐calibrate the scale, follow the instructions outlined below.

1.The patient and the sling MUST be removed from the scale to properly calibrate the Jasmine Scale. For removing the patient instructions refer to Operation on page 12.

For sling detachment instructions refer to Lifting the Patient on page 14.

2.With the Jasmine Scale on, remove the four screws on the back of the enclosure and remove the front cover to expose the PC board.

3.Press the CAL button located on the PC board. The CAL switch is not labeled but is the only button on the PC board located in the lower right corner. Once the CAL button is pushed ʺCAL1ʺ will be seen in the display window.

NOTE: CAL1 selects the calibration mode using 50 pounds of calibrated weight. CAL2 selects the calibration mode for use with 200 pounds of calibrated weight. Pressing the UNITS key toggles between CAL1 and CAL2.

4.When the desired calibration mode is displayed, press the LOCK/UNLOCK key. NOTE: The display window will now show UnLd.

5.Ensure that there is no load on the scale and press the ZERO key.

6.The display window will show a dashed line (‐‐‐‐‐‐) scrolling across and then the word LOAD will appear in the display window.

7.Perform one (1) of the following:

A.FOR CAL1 OPTION ‐ support 50 pounds of calibrated weight from the scale and press the ZERO key.

B.FOR CAL2 OPTION ‐ support 200 pounds of calibrated weight from the scale and press the ZERO key.

8.The display window will show a dashed line (‐‐‐‐‐‐) scrolling across and then the word DONE will appear in the display window.

9.Remove the weight from the scale and press the LOCK/UNLOCK key.

NOTE: The unit will now be in a temporary test mode and will not lock the weight display. This will allow weight to be loaded and unloaded to check the calibration.

10.Turn the unit off by pressing the ON/OFF key.

11.Turn the unit on by pressing the ON/OFF key. NOTE: Scale will now be in normal operation.

Troubleshooting

SYMPTOM

PROBABLE CAUSE

SOLUTION

Unit does NOT work properly.

Battery failure.

Check battery. Replace if necessary.

 

 

 

Battery has been replaced and unit still

 

Contact Invacare for Service at

does NOT work properly.

 

1-800-333-6900

 

 

 

Display Codes

CAL

CALIBRATION REQUIRED ‐ Indicates improper stored calibration data, calibration is necessary.

OCAP

OVER CAPACITY ‐ Indicates a weight exceeding the capacity has been loaded on the scale.

Part No 1150704

31

Jasmine™ Patient Lift

Image 31
Contents Jasmine Patient Lift Symbol Legend Accessories WarningDate of Manufacture Table of Contents Transferring the Patient Troubleshooting MaintenanceAccessories Limited WarrantySpecial Notes Maintenance Must be performed only by qualified personnelSling Sizes Label LocationJasmine Patient Lift Product ParametersJasmine Scale JLS5 Assembling the Lift Weight LimitationUsing the Sling General GuidelinesLifting the Patient Performing MaintenanceTransferring the Patient Assembling the Mast Assembly to the Base AssemblyAssembling the Patient Lift May occur Attaching the Battery Charger Mounting Bracket to the WallAttaching the Battery Charger to Mast Assembly Ing battery on the battery charger to confirmOperating the Patient Lift OperationActivating a Mechanical Emergency Release Charging the BatteryPreparing the Lift for Use Lifting the PatientPositioning the Lift for Use Attaching a Sling to the Swivel Bar Lift the patient. Refer to Lifting/Moving the Patient Lifting/Moving the PatientLifting/Moving the Patient Transferring to a Commode Chair Transferring the PatientTransferring to a Wheelchair Transferring to a Standard CommodeSymptoms Faults Solution TroubleshootingMaintenance Safety Inspection Checklist MaintenanceLubricating the Lift Cleaning the Sling and the Lift Replacing the Boom ActuatorDetecting Wear and Damage Replacing the Boom Actuator Patient Lift Replacing the Leg ActuatorsReplacing the Swivel Bar or Optional Cradle Attachment Replacing the Swivel Bar or Optional Cradle AttachmentChecking and Tightening Mast Pivot Checking and Tightening Mast PivotRemoving the Swivel Bar/Cradle Attachment AccessoriesInstalling the Jasmine Scale Installing the Swivel Bar or Cradle AttachmentKeypad Functions KEY Indicator Displayed Indicator Location DefinitionOperating the Scale Weighing the PatientReplacing the Battery Replacing the BatteryDisplay Codes TroubleshootingSymptom Probable Cause Solution Calibrating the Jasmine ScaleAccessories Accessories Limited Warranty Lève-personne Jasmine Légende DE Symboles Mise EN Garde Sappliquant AUX AccessoiresMise EN Garde Fonctionnement Table DES MatièresDirectives Générales LE MontageDépannage Entretien AccessoiresGarantie Limitée Seule une personne qualifiée Doit exécuter lentretien Remarques SpécialesAvertissement Emplacement DES ÉtiquettesLève-personne Jasmine Paramètres DU ProduitPèse-personne Jasmine JLS5 Lutilisation de la toile Directives GénéralesAssemblage du lève-personne Poids maximumLentretien Lutilisation du lève-personneLe soulèvement du patient Le transfert du patientLe montage du lève-personne LE MontageLe montage du mât à la base La fixation du chargeur de pile au montage du mât La fixation du support de chargeur de pile au murLe fonctionnement du lève-personne FonctionnementLe soulèvement/labaissement du lève-personne Lactivation du déverrouillage mécanique durgence Le chargement de la batterieLa fixation dune toile à la barre pivotante LE Soulèvement DU PatientLa préparation du lève-personne aux fins dutilisation Le positionnement du lève-personne aux fins dutilisationLa fixation dune toile à la barre pivotante La fixation dune toile à lattachement facultatif du berceauLa fixation dune toile à lattachement facultatif du berceau Le soulèvement/déplacement du patientLe soulèvement/déplacement du patient Le transfert vers une chaise daisance LE Transfert DUN PatientLe transfert vers un fauteuil roulant Le transfert vers une chaise daisance standardDépannage Symptômes Défauts SolutionInstitutionnel EntretienListe de contrôle de l’entretien ArticleNettoyage de la toile et du lève-personne Le remplacement de lactionneur de la ceintureLa lubrification du lève-personne Détection de lusure et de lendommagementLe remplacement de lactionneur de la ceinture VU DU Dessous Le remplacement des actionneurs de jambesRetirez les capuchons latéraux Vérification et serrage du pivot du mât Vérification et serrage du pivot du mâtRetrait de la barre pivotante/lattache du berceau AccessoiresDétail B Attacher LA Barre DE Pivot Linstallation du pèse-personne JasmineLemploi du pèse-personne Affichage DE Emplacement DE Touche Lindicateur DéfinitionLes fonctions du clavier La pesée du patient Le calibrage du pèse-personne Jasmine Le remplacement de la batterieDépannage Symptômes Cause Probable SolutionLes codes daffichage Remarques USA Garantie Limitée