After accessing the Service and Factory Menu,PrintscroCalibl toChartthesubmenu and start printing the ch
The chart contains a complete description of the following procedure:
1.Remove the top portion of the chart by cutting along the dotted line.
2.Enter theChange Stored Parametersubmenu and select parameter 200.
3.Write down the existing value at parameter 200, then enter a new value of 1500.
4.Press thefront panel Resolbuttion until FINE is displayed on the front panel display.
5.Using the front panel menu, set the Copy Reduction to 100%.
6.Load the chart (with the top portion removed at the dotted line) into the automatic docum copy. Make sure the chart is loaded top end first with the print side facing down.
7.Look at the copy and find the
8.Determine the Calibration Value corresponding to this tick mark by reading the “How to Rea sample chart following this procedure. The normal range is 550 to 650 on the scale.
9.Reenter theService and FactoryMenu.
10.Scroll to tChange Stored Parametersubmenu and access parameter 200.
11.Enter the Calibration Value determined from the “How to Read Scale” into parameter loc
12.The calibration test is complete. Check that the calibration is correct by making ano previously used (with the top portion removed). Ideally,
A sample calibration chart is provided on the next page. Refer to it facilitate performing t the calibration value.
Calibration and