3M™ Rapid Detection ReaderManaging Settings

Settings: View Tabs

Tab

Screen

LQC continued

Max Failures Before Lockout

Purpose: Define the number of LQC failures allowed before the assay is Locked Out when LQC Timer is set in RCU. LQC Failure count is reset upon successful completion of LQC. Can be set from 1-9.

Default:

9.

Control ID

 

Purpose: Define control IDs.

Default: Min length 1, Max length 18.

 

 

Timer

Figure 23. Settings: View Timer

 

Tab

The ability to edit the settings of the TIMER tab varies by item and is defined below.

Timer Intervals

Menu Idle Timeout

Edit on CM. Cannot edit through RCU.

Purpose: Define the amount of idle time before the 3M Rapid

Detection Reader returns to the Home screen.

Range: 1 -10 minutes.

Default: 5 minutes.

Update/Upload Start Date/Time

Edit through RCU only.

The date and time used to calculate the start of the following intervals.

Update Interval

Edit through RCU only.

Purpose: Interval at which the Reader checks for changes to settings in a networked environment. Disable by entering 0.

Range: 0 – 1440 minutes.

Default: 1.

Upload Results Interval

Edit through RCU only.

Purpose: Interval at which the Reader sends changes or additions to the result and event logs in a networked environment. Disable by entering 0.

Range: 0 - 1440 minutes.

Default: 0 (Disabled).