3M™ Rapid Detection ReaderTroubleshooting

Test Run Messages – indicated by red port icon

 

Port Display

 

Error Message

 

Cause (s)

 

 

Action

 

 

Fail

 

Sample Error 2

 

Sample flow has not been

 

 

The Test Cartridge is

 

 

 

 

 

 

detected.

 

 

ejected.

 

 

 

 

 

 

Insufficient sample added to the

 

 

Remove and dispose of

 

 

 

 

 

 

Test Cartridge.

 

 

appropriately.

 

 

 

 

 

 

Clogged sample pad.

 

 

Repeat the assay with a

 

 

 

 

 

 

Failure to use supplied assay tip.

 

 

fresh set of components.

 

 

 

 

 

 

 

 

If problem persists

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Insufficient mixing of sample.

 

 

contact Technical

 

 

 

 

 

 

Interfering substances in the

 

 

Support to report the

 

 

 

 

 

 

 

 

problem.

 

 

 

 

 

 

sample.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Analyte levels significantly higher

 

 

 

 

 

 

 

 

 

than the upper reportable range.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Fail

 

Temperature Error

 

TM not at proper temperature to

 

 

Contact Technical

 

 

 

 

 

 

carry out specific assay

 

 

Support to report the

 

 

 

 

 

 

requested.

 

 

problem.

 

 

 

 

 

 

 

 

 

 

 

 

Error XX

 

Port Disabled

 

The identified port is disabled

 

 

Contact Technical

 

 

e.g., Error 2B

 

 

 

due to failing IQC or a transport

 

 

Support to report the

 

 

 

 

 

carriage failure. LED remains

 

 

problem.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

illuminated red.

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Test Run Messages

Message / Symptom

Insertion Error Remove & Reinsert

Followed by:

Timed Out

Cause

The Test Cartridge was not inserted properly into the 3M Rapid Detection Reader or the operator did not wait until prompted.

User has not performed the action required within time period allowed.

Action

Cancel the assay or reinsert the Test Cartridge. Fully insert Test Cartridge until firm resistance is felt. Do not use excessive force. The TM will guide the Test Cartridge into the port.

NOTE: The Test Cartridge must be reinserted into the TM port within approximately 30 seconds or an error message will be displayed and the assay will be aborted.

Repeat assay using a fresh set of components.

40

Page 45
Image 45
3M #3968, #3960 manual Insertion Error Remove & Reinsert Followed by Timed Out