3M™ Rapid Detection ReaderManaging Settings

Settings: Edit Tabs

Tab

Screen

Date/Time

Figure 30. Settings: Edit

 

Date/Time Tab

Edit on CM. Can only edit through RCU if NTP is enabled.

Purpose: Set the current date and time for the 3M Rapid Detection Reader.

If in Auto transfer Mode, the 3M Rapid Detection Reader date and time may be synchronized with the NTP server. The date format in the Reader is variable.

To Change:

1.Use drop down menu to scroll through times zones and select correct time zone offset for your location.

2.Enter Current time.

NOTE: When adjusting for changes in daylight savings time

 

 

 

ensure the time zone offset is adjusted – not only the current

 

 

 

time.

 

 

 

NOTE: After changing the date or time, the CM power must be

 

 

 

cycled to reset auto-IQC and data transfer timers.

 

 

 

NOTE: The date format in the RCU is the same as for other

 

 

 

applications running on that PC.

 

 

 

NOTE: Periodically check that the clock is correct and has not

 

 

 

been inadvertently changed.

 

 

 

 

 

 

 

Language

Figure 31.

Settings: Edit

 

Edit on CM only.

Language Tab

 

 

 

 

 

 

 

Purpose: Set the display language for the 3M Rapid Detection

 

 

 

Reader.

 

 

 

 

 

 

 

 

 

 

 

Misc.

Figure 32.

Settings: Edit Misc. Tab

 

Manual / Auto Transfer Mode

 

 

 

Edit on CM. Cannot edit through RCU.

 

 

 

Manual Transfer Mode: 3M Rapid Detection Reader is set up for

 

 

 

stand-alone operation.

 

 

 

Auto Transfer Mode: 3M Rapid Detection Reader is set up for

 

 

 

network operation.

 

 

 

Default: Manual Transfer Mode.

 

 

 

 

 

 

25

Page 30
Image 30
3M #3960, #3968 manual Date/Time, Language, Misc, Manual / Auto Transfer Mode