A&D HV-G Series instruction manual Comparator, Full/Dribble batch sub-function, Hold

Models: HV-G Series

1 69
Download 69 pages 38.25 Kb
Page 34
Image 34

Caution

The comparison of the full/dribble batch function is not reversible (One way se- quence). If the display value becomes less than the final value after the value reached a predetermined target value, neither HI or LO is on.

The parameters of the upper limit value (HI) and the final value (HI) use the same memory. The parameters of the lower limit value (LO) and the zero band (LO) use the same memory.

The upper/lower comparator function, the simple batch function and the full/dribble batch function cannot be used at the same time because these parameters use common memory.

Set the zero band greater than the tare value.

Operation

Pressing the SET switch, the scale starts the batch process.

Selecting a parameter from f10 0 or f10 2 of the full/dribble batch sub-function, the F switch works as the finish switch.

Parameter List and Word Definition

The “gross” is a total weighing value, no tare value is subtracted.

The “net” is a measurement value with the tare value subtracted from the gross. The “zero band” means the zero detection level.

The “zero point” is the fundamental starting point to weighing.

Comparator

Function tableMeaning and purpose

f6 9 Full/Dribble batch function

Full/Dribble batch sub-function

 

Function table

 

Meaning and purpose

 

 

f10

0

 

Reaching final value and pressing the F switch the current process is

 

 

 

 

 

finished.

 

 

 

f10

1

 

Reaching the final value and displaying the stability mark the current

 

 

 

 

 

process is finished automatically.

 

 

 

f10

2

 

Pressing the

 

 

switch the scale automatically tares and starts the

 

 

 

 

SET

 

 

 

 

 

full/dribble batch process. Reaching final value and pressing the F switch,

 

 

 

 

 

the current process is finished.

 

 

f10

3

 

Pressing the

 

switch the scale automatically tares and starts the

 

 

 

 

SET

 

 

 

 

 

 

full/dribble batch process. Reaching the final value and displaying the

 

 

 

 

 

stability mark the current process is finished.

Hold

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Function table

 

Meaning and purpose

 

 

f12

0

 

The hold function is not used

 

 

 

 

 

 

 

 

 

 

 

 

 

HV-G Instruction Manual Page 33

Page 34
Image 34