A&D HV-G Series instruction manual Switches

Models: HV-G Series

1 69
Download 69 pages 38.25 Kb
Page 15
Image 15

Calibration error.

Means “Calibration mass is too heavy”.

Check the base unit and weighing pan.

Weighed value is unstable due to drift, vibration etc. when turning on the scale.

Check around the weighing pan. Check the connection of load cell cable.

Remove everything from the weighing pan.

Check around weighing pan.

Perform zero point calibration of the scale.

Accumulated data count

Total mass value of the accumulated data.

Comparator function = display is an upper limit.

Full/dribble batch function = display is a final value.

Full/dribble batch function = display is a preliminary value.

Comparator function = display is a lower limit.

Full/dribble batch function = display is the zero band.

Description of the weighing unit, weighing range and minimum graduation.

6.2 Switches

Power switch.

Note Type V is in standby status when power is connected, even though the display is off.

Zero switch.

When there is nothing on the pan and the ZERO switch is pressed, the scale displays zero and the zero point mark. Any stored tare is cancelled.

Tare switch.

Used to store the weight of a container when carrying out net weighing.

Range switch.

For changing the weighing range. (Refer to f2 )

Storing the unit mass, it is used to select a sample number.

In the function table, it is used to select a parameter.

HV-G Instruction Manual Page 14

Page 15
Image 15
A&D HV-G Series instruction manual Switches