Upper limit value (HI) Lower limit value (LO)

11.2 Operation and Performance (Examples)

Example 1

This example is set as follows:

Function table f6 3 (If the current display value is not near-zero, the scale compares the display value with the

upper limit value and the lower limit value immediately.)

7.000kg

6.500kg

Case

The comparison starts at turning the scale on.

When the current value is less than 6.500kg, LO is displayed.

When the current value is between 6.500kg and 7.000kg, OK is displayed. When the current value is greater than 7.000kg, HI is displayed.

Example 2

This example is set as follows:

 

 

 

 

Function table

 

(Pressing the

 

 

switch, after the stability

f6 4

SET

 

 

mark is displayed, the scale compares the

 

 

current display value with the upper limit value

 

 

and the lower limit value immediately.)

Upper limit value (HI)

2.000kg

Lower limit value (LO)

-1.000kg

Case

Pressing the SET stability mark.

switch, the comparison is performed after displaying the

When the current value is less than -1.000kg, LO is displayed.

When the current value is between -1.000kg and 2.000kg, OK is displayed.

When the current value is greater than 2.000kg, HI is displayed.

HV-G Instruction Manual Page 31

Page 32
Image 32