Item

Display

Meaning and purpose

 

 

 

 

 

 

 

 

 

 

 

Accumulation

 

When the display becomes a stable value, the scale

 

 

accumulates the data automatically. Next accumulation can

 

 

perform after the display becomes near-zero.

 

f8 4

Use: Recording number and mass of articles removed from

 

 

the pan. (Put articles on the pan. Press

 

switch at each

 

 

TARE

 

 

removal.)

 

 

At each finish of full/dribble batch function, the scale

 

 

accumulates the data automatically.

 

f8 5

Use: Packing articles like a powder, it is used for recording

 

 

the bag number and total mass.

 

 

 

 

The printing condition when using option OP-06 for type V.

 

f9 0 #

No print.

 

f9 1

Pressing the

 

 

switch, the data is output.

 

PRINT

 

 

Auto-print +

 

 

When the display becomes a positive stable value above

 

f9 2

near-zero, the scale outputs the data automatically. Next

 

 

output can be performed after the display becomes near-

 

 

zero or negative value.

Print Mode

 

Auto-print +/-

 

 

When the display becomes a stable value outside near-zero,

 

f9 3

the scale outputs the data automatically. Next output can be

 

 

performed after the display becomes near-zero.

 

 

 

 

 

 

 

 

 

 

 

 

f9 4

At each finish of full/dribble batch function, the data is output.

 

f9 5

When the accumulation function is used, the data is output.

 

 

The details of the full/dribble batch function (B$￿')

 

f10 0 #

Reaching final value and pressing the F switch, the current

 

 

process is finished.

 

f10 1

Reaching final value and displaying the stability mark, the

 

 

current process is finished automatically.

 

f10 2

Pressing the

 

 

switch, the scale automatically tares and

 

 

SET

 

 

starts the full/dribble batch process. Reaching final value and

 

 

pressing the F switch, the current process is finished.

Full/Dribble

f10 3Pressing the

 

switch, the scale automatically tares and

SET

Batch

starts the full/dribble batch process. Reaching final value and

sub-function

displaying the stability mark, the current process is finished.

 

Address for RS-422 / RS-485 for option OP-04

 

f11 00 # Use This address can be set from 01 to 99. Allows a

Address

to

computer to control a scale with the assigned address.

 

f11 99

(RS-232C should be set to “00”.)

 

 

 

 

 

 

 

 

 

 

 

HV-G Instruction Manual Page 46

Page 47
Image 47
A&D HV-G Series instruction manual Printing condition when using option OP-06 for type