106 6000 Series Oscilloscope User’s Guide
2 Front-Panel Controls
Performing User Cal will invalidate your Certificate of
Calibration. If NIST (National Institute of Standards and
Technology) traceability is required, perform the Performance
Verification procedure in the Agilent InfiniiVision 6000 Series
Oscilloscopes Service Guide using traceable sources.
To perform User Cal
1Set the rear-panel CALIBRATION switch to UNPROTECTED.
2Connect short (12 inch maximum) equal length cables to
each analog channel’s BNC connector on the front of the
oscilloscope. You will need two equal-length cables for a
2-channel oscilloscope or four equal-length cables for a
4-channel oscilloscope.
Use 50 RG58AU or equivalent BNC cables when performing
User Cal.
For a 2-channel oscilloscope, connect a BNC tee to the equal
length cables. Then connect a BNC(f)-to-BNC(f) (also called a
barrel connector) to the tee as shown below.
Figure 10 User Calibration cable for 2-channel oscilloscope
Longer cable
to TRIG OUT
To Channel 1
To Channel 2