Triggering the Oscilloscope 4
6000 Series Oscilloscope User’s Guide 151
5Repeatedly press and release the Baud softkey to set the CAN
signal baud rate to match your CAN bus signal.
The CAN baud rate can be set to:
The default baud rate is 1 Mb/s
6Repeatedly press and release the Smpl Pt softkey to select the
point between phase segments 1 and 2 where the state of the
bus is measured. This controls the point within the bit’s time
where the bit value is captured.
You ca n us e th e Zoom mode for easier navigation of the decoded
data.
10 kb/s 50 kb/s 100 kb/s 500 kb/s
20 kb/s 62.5 kb/s 125 kb/s 800 kb/s
33.3 kb/s 83.3 kb/s 250 kb/s 1 Mb/s
One Bit
60%
70%
80%
Sample Point
NOTE If the setup does not produce a stable trigger, the CAN signal may be slow
enough that the oscilloscope is AutoTriggering. Press the Mode/Coupling
key, then press the Mode softkey to set the trigger mode from Auto to
Normal.