194 6000 Series Oscilloscope User’s Guide
4 Triggering the Oscilloscope
To trigger on a specific field of the video signal
To examine the components of a video signal, trigger on either
Field 1 or Field 2 (available for interleaved standards). When a
specific field is selected, the oscilloscope triggers on the rising
edge of the first serration pulse in the vertical sync interval in
the specified field (1 or 2).
1Press the Trigger More key, then press the TV softkey.
2Press the Settings softkey, then press the Standard softkey to
select the appropriate TV standard.
3Press the Mode softkey and select Field1 or Field2.
Figure 17 Triggering on Field 1