34 6000 Series Oscilloscope User’s Guide
1Getting Started
The 6000A Series Option BAT oscilloscopes are battery
powered. They have additional LED indicators on the front
panel:
Battery life is approximately 1.75 hours, depending on the
oscilloscope configuration.

Operating with the Internal Battery

When operating with the internal battery, the operating
temperature should be in the range from -10°C to 50°C, 80%
relative humidity.
1Make sure battery is charged before use. See “Charging the
Battery” below.
2Connect the supplied ground wire from the ground post on
the back of the instrument to earth ground.
Caution indicator. Lights (amber) when running
on the internal battery. See “Operating with the
Internal Battery” below.
Battery power indicator. Turns from green to red
when there is 15 to 20 minutes of battery power
remaining.
Charging indicator. Lights when the battery is
charging. Turns off when the battery is fully
charged or when the charger is disconnected.
WARNING Maintain oscilloscope ground connection
Do not negate the protective action of the ground connection to the
oscilloscope. If the circuit under test has voltages greater than 30
Vrms, the oscilloscope must be grounded through its ground
terminal to prevent electric shock to the user.