144 6000 Series Oscilloscope User’s Guide
4 Triggering the Oscilloscope
For example, for a positive pulse, if you set t>10ns:
within a range of time values (><).
For example, for a positive pulse, if you set t>10ns and
t<15 ns:
5Select the qualifier time set softkey (< or >), then rotate the
Entry knob to set the pulse width qualifier time.
The qualifiers can be set as follows:
2 ns to 10 s for > or < qualifier (5 ns to 10 s for 100 MHz
and 300 MHz bandwidth models)
10 ns to 10 s for >< qualifier, with minimum difference of
5 ns between upper and lower settings
< qualifier time set softkey
When the less than (<) qualifier is selected, the Entry knob
sets the oscilloscope to trigger on a pulse width less than the
time value displayed on the softkey.
When the time range (><) qualifier is selected, the Entry
knob sets the upper time range value.
> qualifier time set softkey
When the greater than (>) qualifier is selected, the Entry
knob sets the oscilloscope to trigger on a pulse width greater
than the time value displayed on the softkey.
When the time range (><) qualifier is selected, the Entry
knob sets the lower time range value.
Trigger
10ns10ns
Trigger
12ns
15ns
10ns