108 6000 Series Oscilloscope User’s Guide
2 Front-Panel Controls
probes do not need to be calibrated, but InfiniiMax probes can
be calibrated. For more information about calibrating probes
see page 55.
Results:
User Cal date:
Change in temperature since last User Cal:
Failure:
Comments:
Probe Cal Status:
Self Test
Pressing Utility&Service&Start Self Test performs a series of
internal procedures to verify that the oscilloscope is operating
properly.
It is recommended you run Self Test:
after experiencing abnormal operation.
for additional information to better describe an oscilloscope
failure.
to verify proper operation after the oscilloscope has been
repaired.
Successfully passing Self Test does not guarantee 100% of the
oscilloscope's functionality. Self Test is designed to provide an
80% confidence level that the oscilloscope is operating properly.
About Oscilloscope
Pressing Utility&Service&About Oscilloscope displays information
about your oscilloscope model number, serial number, software
version, boot version, graphics version, and installed licenses.