160 6000 Series Oscilloscope User’s Guide
4 Triggering the Oscilloscope
Triggering on FlexRay Frames, Times, or Errors

Accessing the FlexRay Trigger Menu

1In the Trigger section on the front panel, press the More key.
2Press the Trigger softkey, and use the Entry knob to select
FlexRay.
3Press the Settings softkey to access the FlexRay Trigger Menu.

Triggering on FlexRay Frames

1In the FlexRay Trigger Menu (see “Accessing the FlexRay
Trigger Menu”on page 160), press the Trigger softkey, and use
the Entry knob to select Frame.
2Press the Frames softkey to access the FlexRay Frame Trigger
Menu.
3Press the CycCt Rep sof tkey, and use the Entry knob to select
the cycle count repeat value from 2, 4, 8, 16, 32, or 64.
4Press the CycCt B as softkey, and use the Entry knob to select
the cycle count base value from 0 through the CycCt Rep value
minus 1, or All. For example, with a base value of 1 and a
repetition value of 16, the oscilloscope triggers on cycles 1,
17, 33, 49, and 65.
Trigger
condition
VPT1000 Menu
FlexRay
trigger
Return to
previous menu
Frames, Time, Errors Selection