Triggering the Oscilloscope 4
6000 Series Oscilloscope User’s Guide 195
To trigger on all fields of the video signal
To quickly and easily view transitions between fields, or to find
the amplitude differences between the fields, use the All Fields
trigger mode.
1Press the Trigger More key, then press the TV softkey.
2Press the Settings softkey, then press the Standard softkey to
select the appropriate TV standard.
3Press the Mode softkey and select All Fields.
Figure 18 Triggering on All Fields