180 6000 Series Oscilloscope User’s Guide
4 Triggering the Oscilloscope
When Nth Edge 2 (no re-find) is selected, if the Find event
occurs again before the Count (N) event is satisfied, the
Count (N) is not reset to zero.
bPress the Term softkey and select Count (N).
cPress the N softkey, then turn the Entry knob to set the
number of edges to wait before triggering.
N can be set from 1 to 10,000.
Define the optional Reset on: stage1Press the Stage softkey and select Reset on:.
Reset on: is the last stage in the trigger sequence. When you
select the Stage Reset on: softkey, the next softkey to the right
will display Reset: and give you a list of the terms you can
define for the Reset on: stage. The Reset on: stage can be set
to one of the following conditions:
•No Reset — Resets on the find condition.
•Pattern 1 (or 2) Entered — A pattern is entered on the last
edge that makes the pattern true (logical AND).
•Pattern 1 (or 2) Exited — A pattern is exited on the first edge
that makes the pattern false (logic NAND).
•Edge 1 (or 2)
•Pattern 1 and Edge 1
•Timeout
Terms that are grayed-out are not valid in the reset stage.
2Press the Reset: softkey and select a term to reset on.
3Press the Term softkey and select the pattern, edge, or timeout
term displayed in the Reset: softkey.
4If you select No Reset, no reset stage will be defined.
If the term(s) used in the “Trigger on:” Sequence Stage condition are set to “don’t
care,” the oscilloscope will not trigger. At least one term in the stage must be set
to a value other than X (don’t care).