4Press the CycCt B as softkey, and use the Entry knob to select
the cycle count base value from 0 through the CycCt Rep value
minus 1, or All.
5Press the Segment softkey to select the type of segments from:
Static Segments
Dynamic Segments
Symbol Segments
Idle Segments
6When Static or Dynamic Segments are selected:
aPress the Slot# softkey, and use the Entry knob to select
the time slot number value from 1 to 2047 or All. The
maximum static slot value and the minimum dynamic slot
value are limited by the downloaded FIBEX file schedule.
bPress the SlotType softkey to select the slot type:
All Slots — the oscilloscope triggers on slots regardless of
whether they contain a frame.
Empty Slots — the oscilloscope triggers when a slot does
not contain a frame. Usually used to determine if an
assigned slot is empty when it should have contained a
frame.

Triggering on FlexRay Errors

1In the FlexRay Trigger Menu (see “Accessing the FlexRay
Trigger Menu”on page 160), press the Trigger softkey, and use
the Entry knob to select Error.
2Press the Errors softkey, and use the Entry knob to select the
error type from:
All Errors
Code Error (NRZ) — NRZ decoding errors only.
NOTE Since up to 64 cycles can be defined for FlexRay it may be helpful to set the
oscilloscope to Normal trigger mode instead of Auto trigger mode. This
prevents the oscilloscope from Auto triggering while waiting for a
particular slot in a particular cycle.