170 Chapter 2
Instrument Functions: A - L
FREQUENCY / Channel
Instrument Functions: A - L
Dependencies/
Couplings: Signal Track is turned off when you do any of the following:
1. Preset is pressed (with Preset Type set to Factory),
2. Markers All Off is pressed.
Signal Track is not available when any of the following is true:
1. Signal ID = on
2. Span = 0 Hz
3. Frequency scale type = Log
4. Segmented sweep = on
The following features are not available when Signal Track is on:
1. Signal ID
2. Segmented sweep
3. Log sweep
State Saved: Saved in instrument state.
Factory Preset: Off
Remote Command:
:CALCulate:MARKer[1]|2|3|4:TRCKing[:STATe] OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4:TRCKing[:STATe]?
Example: CALC:MARK1:TRCK ON turns on Signal Track using Marker 1.
CALC:MARK1:TRCK?.