Chapter 5 351
One-Button Measurement Functions
Meas Setup (Adjacent Channel Power—ACP)
One-Button Measurement Functions

5.6 Meas Setup (Adjacent Channel Power—ACP)

If the adjacent channel power (ACP) measurement has been selected in the Measure menu
of the spectrum analysis mode, this key displays the ACP measurement setup menu.
The adjacent channel power measurement measures the power that leaks into adjacent
transmit channels. Depending on the radio standard chosen from the Mode Setup menu,
this measurement can run in several different modes in order to measure different types of
signals.
Key Path: Front-panel key
History: Added with firmware revision A.02.00.
Remote Command:
There is no equivalent remote command.

5.6.1 Avg Number

Pressing Avg Number to On enables you to specify the number of measurements that will be
averaged when calculating the measurement result. The average will be displayed at the
end of each sweep.Setting Avg Number to Off disables the measurement averaging.
Key Path: Meas Setup
State Saved: Saved in instrument state.
Factory Preset: 10 / Off
Range: 1 to 1000
History: Added with firmware revision A.02.00.
Remote Command:
[:SENSe]:ACPower:AVERage:COUNt <integer>
[:SENSe]:ACPower:AVERage:COUNt?
[:SENSe]:ACPower:AVERage[:STATe] OFF|ON|0|1 turns the averaging on or off.
[:SENSe]:ACPower:AVERage[:STATe]?
Example: ACP:AVER:COUN 10
ACP:AVER:COUN?
ACP:AVER OFF
ACP:AVER?